312 Chapter 2 
Performance Verification Tests
36. Displayed Average Noise Level: Agilent E4402B and E4403B
Figure 2-54.
17.Disconnect the BNC cable and adapter from the AMPTD REF OUT 
and the 50 Ω Input.
18.If the analyzer has Option 1DN, 50 Ω tracking generator, do the 
following:
a. On the analyzer, press BW/Avg, Res BW, 1 kHz.
b. Press Source, Amplitude, 0 dBm.
c. Connect a 50 Ω termination to the RF OUT 50 Ω.
Measurement Sequence
The following DANL Measurement Sequence table lists the procedures 
to be performed and the parameters to be used in each procedure. Also 
listed in the table are test record entry numbers for recording the 
results in the performance verification test record.
1. Perform all of the following steps (through step 7) that apply to your 
analyzer using the appropriate subsets in Table 2-85 (for E4402B) or 
Table 2-86 (for E4403B). Then record the display line amplitude 
setting as the indicated Test Record entry in the performance 
verification test record.
2. If the minimum RBW of the analyzer is 1 kHz, perform those 
procedures listed as Subset A in Table 2-85 or Table 2-86.
3. If the minimum RBW of the analyzer is 1 kHz and Option 1DS (RF 
Preamplifier) is installed, also perform those procedures listed in 
Subset B in Table 2-85.
4. If the minimum RBW of the analyzer is 100 Hz or 10 Hz and Option 
1DN (3.0 GHz Tracking Generator) is installed, perform those 
procedures listed in Subset A in Table 2-85 or Table 2-86.
5. If the minimum RBW of the analyzer is 10 Hz and both Option 1DS 
(RF Preamplifier) and Option 1DN (3.0 GHz Tracking Generator) 
are installed, also perform those procedures listed in Subset B in 
Table 2-85.
6. If the minimum RBW of the analyzer is 100 Hz or 10 Hz, perform 
those procedures listed in Subset C in Table 2-85 or Table 2-86.
7. If the minimum RBW of the analyzer is 10 Hz and Option 1DS (RF 
Preamplifier) is installed, also perform those procedures listed in