588 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
10 MHz to 1 GHz (18)________ –150 dBm ±1.23 dB
1 GHz to 2 GHz (19)________ –150 dBm ±1.23 dB
2 GHz to 3 GHz (20)________ –146 dBm ±1.23 dB
1kHz RBW, Preamp On,
20 to 30° C
10 MHz to 1 GHz (21)________ –132 dBm ±1.23 dB
1 GHz to 2 GHz (22)________ –132 dBm ±1.23 dB
2 GHz to 3 GHz (23)________ –130 dBm ±1.23 dB
10 Hz RBW, Preamp On,
20 to 30° C
10 MHz to 1 GHz (24)________ –151 dBm ±1.23 dB
1 GHz to 2 GHz (25)________ –151 dBm ±1.23 dB
2 GHz to 3 GHz (26)________ –149 dBm ±1.23 dB
39. Residual Responses
150 kHz to 6.7 GHz (1)________ –90 dBm ±0.87 dB
40. Fast Time Domain
Amplitude Accuracy
(Option AYX only)
Amplitude Error –0.3% (1)________ 0.3% ±0.029%
42. Tracking Generator
Absolute Amplitude
and Vernier Accuracy
(Option 1DN only)
Absolute Amplitude
Accuracy
–0.75 dB (1)________ 0.75 dB ±0.096 dB
Absolute Vernier
Accuracy
–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB
–3 dB –0.5 dB (3)________ 0.5 dB ±0.008 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty