Chapter 3 581
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Minimum Response –3.0 dB (32)________ ±0.40 dB
Peak-to-Peak Response (33)________ 4.0 dB ±0.40 dB
ac Coupled
Band 3
(13.2 GHz to 26.5 GHz)
Maximum Response (34)________ 3.0 dB ±0.59 dB
Minimum Response –3.0 dB (35)________ ±0.59 dB
Peak-to-Peak Response (36)________ 4.0 dB ±0.59 dB
27. Frequency Response
(Preamp On)
(Option 1DS Only)
Note: Enter data in the appropriate section below depending upon the
ambient temperature at which the test was performed and installed
options.
20 to 30° C
Non-Option UKB
dc Coupled
Band 0
(1 MHz to 3.0 GHz)
Maximum Response (1)________ 1.5 dB ±0.28 dB
Minimum Response –1.5 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 3.0 dB ±0.28 dB
Non-Option UKB
dc Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (4)________ 1.5 dB ±0.28 dB
Minimum Response –1.5 dB (5)________ ±0.28 dB
Peak-to-Peak Response (6)________ 3.0 dB ±0.28 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty