Chapter 2 223
Performance Verification Tests
26. Frequency Response (Preamp On): Agilent E4402B
15.Set the synthesized sweeper CW frequency to 10 MHz and the
amplitude to 6 dBm.
16.Adjust the synthesized sweeper power level to obtain a Channel A
power meter reading of –20 dBm ±0.1 dB.
17.Record both Channel A and Channel B power meter readings in
Table 2-60.
18.Repeat step 9 through step 12 for each remaining frequency in
Table 2-60.
19.For each entry in Table 2-60, calculate the Splitter Tracking Error as
follows:
For example, if Channel A Power is –20.3 dBm and Channel B power
is –0.23 dBm, the splitter tracking error is –20.07 dB.
Splitter Tracking Error Channel A Power Channel B Power=
Table 2-60 Source/Splitter Characterization
Frequency
Power Meter Reading Splitter
Tracking
ErrorChannel A Channel B
1 MHz
5 MHz
10 MHza
10 MHzb
20 MHz
50 MHz
75 MHz
175 MHz
275 MHz
375 MHz
475 MHz
575 MHz
675 MHz
775 MHz
825 MHz
875 MHz
925 MHz