Chapter 3 623
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
–84 –1.14 dB (63)________ 1.14 dB ±0.086 dB
–88 –1.18 dB (64)________ 1.18 dB ±0.085 dB
–92 –1.22 dB (65)________ 1.22 dB ±0.100 dB
–96 –1.26 dB (66)________ 1.26 dB ±0.099 dB
–98 –1.28 dB (67)________ 1.28 dB ±0.098 dB
Incremental Log Fidelity,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.4 dB (68)________ 0.4 dB ±0.082 dB
–8 –0.4 dB (69)________ 0.4 dB ±0.078 dB
–12 –0.4 dB (70)________ 0.4 dB ±0.075 dB
–16 –0.4 dB (71)________ 0.4 dB ±0.073 dB
–20 –0.4 dB (72)________ 0.4 dB ±0.078 dB
–24 –0.4 dB (73)________ 0.4 dB ±0.074 dB
–28 –0.4 dB (74)________ 0.4 dB ±0.073 dB
–32 –0.4 dB (75)________ 0.4 dB ±0.077 dB
–36 –0.4 dB (76)________ 0.4 dB ±0.075 dB
–40 –0.4 dB (77)________ 0.4 dB ±0.081 dB
–44 –0.4 dB (78)________ 0.4 dB ±0.077 dB
–48 –0.4 dB (79)________ 0.4 dB ±0.076 dB
–52 –0.4 dB (80)________ 0.4 dB ±0.080 dB
–56 –0.4 dB (81)________ 0.4 dB ±0.078 dB
–60 –0.4 dB (82)________ 0.4 dB ±0.084 dB
–64 –0.4 dB (83)________ 0.4 dB ±0.081 dB
–68 –0.4 dB (84)________ 0.4 dB ±0.080 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty