Chapter 3 633
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
Positive Level Flatness,
>10 MHz
(5)________ 2.0 dB ±0.19 dB
Negative Level Flatness,
>10 MHz
–2.0 dB (6)________ ±0.19 dB
45. Tracking Generator
Harmonic Spurious
Outputs (Option 1DN or
Option 1DQ only)
2nd Harmonic, <20 MHz (1)________ –20 dBc ±2.6 dB
2nd Harmonic, 20 MHz (2)________ –25 dBc ±2.6 dB
3rd Harmonic, <20 MHz (3)________ –20 dBc ±2.6 dB
3rd Harmonic, 20 MHz (4)________ –25 dBc ±2.6 dB
47. Tracking Generator
Non-Harmonic
Spurious Outputs
(Option 1DN or
Option 1DQ only)
Highest Non-Harmonic
Spurious Output
Amplitude
(1)________ –35 dBc ±2.64 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty