Chapter 2 359
Performance Verification Tests
45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B  (Option
1DN or 1DQ)
f. Press MKR, MARKERS (Off).
9. From Table 2-99, enter the 2nd Harmonic Level at 10 MHz as Test 
Record entry 1 and copy this value into the performance verification 
test record.
• Test Record entry 1: 
TG 2nd Harmonic Spurious Output __________ dB
10.From Table 2-99, locate the most positive 2nd Harmonic Level for 
tracking generator frequencies of 100 MHz to 750 MHz and record 
this value as Test Record entry 2 and copy this value into the 
performance verification test record.
• Test Record entry 2: 
TG 2nd Harmonic Spurious Output __________ dB
11.From Table 2-99, enter the 3rd Harmonic Level at 10 MHz as Test 
Record entry 3 and copy this value into the performance verification 
test record.
• Test Record entry 3: 
TG 3rd Harmonic Spurious Output __________ dB
12.From Table 2-99, locate the most positive 3rd Harmonic Level for 
tracking generator frequencies of 100 MHz to 750 MHz and record 
this value as Test Record entry 4 and copy this value into the 
performance verification test record.
• Test Record entry 4: 
TG 3rd Harmonic Spurious Output __________ dB
Table 2-99  Tracking Generator Harmonic Spurious Response Worksheet 
1.5 GHz 
Tracking Generator 
Output Frequency 
2nd Harmonic Level 
(dBc) 
3rd Harmonic Level 
(dBc) 
10 MHz
100 MHz
300 MHz
750 MHz N/A