46. Tracking Generator 
Harmonic Spurious 
Outputs
(Option 1DN only)
2nd Harmonic, <20 kHz (1)________ –15 dBc ±2.6 dB
2nd Harmonic, ≥ 20 kHz (2)________ –25 dBc ±2.6 dB
3rd Harmonic, <20 kHz (3)________ –15 dBc ±2.6 dB
3rd Harmonic, ≥ 20 kHz (4)________ –25 dBc ±2.6 dB
48. Tracking Generator 
Non-Harmonic 
Spurious Outputs 
(Option 1DN only)
  Highest Non-Harmonic 
Spurious Output 
Amplitude, 9 kHz to 2 GHz
(1)________ –27 dBc ±2.63 dB
  Highest Non-Harmonic 
Spurious Output 
Amplitude, 2 GHz to 
3GHz
(2)________ –23 dBc ±3.14 dB
49. Tracking Generator 
L.O. Feedthrough  
(Option 1DN only)
  9 kHz to 2.9 GHz (1)________ –16 dBm ±2.58 dB
  2.9 GHz to 3.0 GHz (2)________ –16 dBm ±3.03 dB
50. Gate Delay Accuracy 
and Gate Length 
Accuracy 
(Option 1D6 only)
Minimum Gate Delay 499.9 ns (1)________ 1.5001 µs±549 ps
Maximum Gate Delay 499.9 ns (2)________ 1.5001 µs±549 ps
1 µs Gate Length 499.9 ns (3)________ 1.5001 µs±520 ps
Table 3-8  Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 