520 Chapter 3
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
46. Tracking Generator
Harmonic Spurious
Outputs
(Option 1DN only)
2nd Harmonic, <20 kHz (1)________ –15 dBc ±2.6 dB
2nd Harmonic, 20 kHz (2)________ –25 dBc ±2.6 dB
3rd Harmonic, <20 kHz (3)________ –15 dBc ±2.6 dB
3rd Harmonic, 20 kHz (4)________ –25 dBc ±2.6 dB
48. Tracking Generator
Non-Harmonic
Spurious Outputs
(Option 1DN only)
Highest Non-Harmonic
Spurious Output
Amplitude, 9 kHz to 2 GHz
(1)________ –27 dBc ±2.63 dB
Highest Non-Harmonic
Spurious Output
Amplitude, 2 GHz to
3GHz
(2)________ –23 dBc ±3.14 dB
49. Tracking Generator
L.O. Feedthrough
(Option 1DN only)
9 kHz to 2.9 GHz (1)________ –16 dBm ±2.58 dB
2.9 GHz to 3.0 GHz (2)________ –16 dBm ±3.03 dB
50. Gate Delay Accuracy
and Gate Length
Accuracy
(Option 1D6 only)
Minimum Gate Delay 499.9 ns (1)________ 1.5001 µs±549 ps
Maximum Gate Delay 499.9 ns (2)________ 1.5001 µs±549 ps
1 µs Gate Length 499.9 ns (3)________ 1.5001 µs±520 ps
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty