Chapter 3 437
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
–56 –1.14 dB (116)________ 1.14 dB ±0.078 dB
–60 –1.20 dB (117)________ 1.20 dB ±0.084 dB
–64 –1.5 dB (118)________ 1.5 dB ±0.081 dB
–68 –1.5 dB (119)________ 1.5 dB ±0.080 dB
–70 –1.5 dB (120)________ 1.5 dB ±0.084 dB
13. Input Attenuation
Switching Uncertainty
Input Attenuation Setting
0 dB –0.3 dB (1)________ 0.3 dB ±0.101 dB
5 dB –0.3 dB (2)________ 0.3 dB ±0.104 dB
15 dB –0.3 dB (3)________ 0.3 dB ±0.102 dB
20 dB –0.3 dB (4)________ 0.3 dB ±0.098 dB
25 dB –0.35 dB (5)________ 0.35 dB ±0.098 dB
30 dB –0.40 dB (6)________ 0.40 dB ±0.096 dB
35 dB –0.45 dB (7)________ 0.45 dB ±0.099 dB
40 dB –0.50 dB (8)________ 0.50 dB ±0.096 dB
45 dB –0.55 dB (9)________ 0.55 dB ±0.099 dB
50 dB –0.60 dB (10)________ 0.60 dB ±0.095 dB
55 dB –0.65 dB (11)________ 0.65 dB ±0.099 dB
60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB
14. Reference Level
Accuracy
Log, Res BW 1 kHz
Reference Level
50 (dBm) 75 (dBmV)
–15 33.75 –0.3 dB (1)________ 0.3 dB ±0.091 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty