120 Chapter 2
Performance Verification Tests
15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
E4408B.

performance test record.

The following is an example for –30 dBm reference level:

Log Scale, Digital Bandwidths, Option 1DR

NOTE If the analyzer is an E4404B/E4405B/E4407B that is not equipped with

Option 1DR (narrow resolution bandwidths) or an E4403B/E4408B

(with or without Option 1DR), skip to the next section (Linear Scale,

Analog Bandwidths).

7. On the analyzer, press the following keys:

SPAN, 150 Hz
BW/Avg, 10 Hz
Video BW, 1 Hz

8. Set the 1 dB step attenuator to set the signal peak 1 to 3 divisions

below the reference level.

Analyzer marker amplitude 10.17() dB=

Actual Attenuator Step 30 dB()10.07() dB=

Test Record Entry 10.17() dB 10.07() dB 0.10() dB==

Table 2-24 Log Mode, Analog Bandwidths Worksheet

Column 1 Column 2 Column 3 Column 4 Column 5 Column 6
10 dB
Attenuator
Nominal
Attenuation
(dB)
10 dB
Attenuator
Actual
Attenuation
(dB)
Actual
Attenuator
Step
(dB)
Analyzer
Reference
Level
(dBm)
Analyzer
Marker
Delta
Amplitude
(dB)
Test
Record
Entry
20 0 (Ref) 20 0 (Ref) (Ref)
10 10 1)
002)
30 30 3)
40 40 4)
50 50 5)
60 60 6)
70 70 7)
80 80 8)