Chapter 2 83
Performance Verification Tests
11. Sweep Time Accuracy
11. Sweep Time Accuracy
This test uses a function generator to amplitude-modulate a 500 MHz 
CW signal from another signal generator. The analyzer demodulates 
this signal in zero span to display the response in the time domain. The 
marker ∆ function on the analyzer is used to read out the sweep time 
accuracy.
If the analyzer is equipped with Option AYX, also perform “Fast Time 
Domain Amplitude Accuracy” in addition to this procedure.
There are no related adjustment procedures for this performance test. 
Equipment Required 
Function generator 
Synthesized signal generator 
Cable, Type-N, 152-cm (60-in) 
Cable, BNC, 120-cm (48-in) 
Additional Equipment for 75 Ω Input
Pad, minimum loss
Adapter, Type-N (f), to BNC (m), 75 Ω
Additional Equipment for Option BAB
Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-11  Sweep Time Accuracy Test Setup 
CAUTION Use only 75 Ω cables, connectors, or adapters on instruments with 75 Ω