Chapter 2 261
Performance Verification Tests
30. Spurious Responses: Agilent E4401B and E4411B
36.Repeat step 10 through step 29. This is the TOI test for Option 1DR.
37.On the analyzer, press System, Alignments, Auto Align, On.
38.Part 1: Third Order Intermodulation Distortion is complete.
Continue with Part 2: Second Harmonic Distortion.
Part 2: Second Harmonic Distortion
1. Set the synthesized signal generator controls as follows:
FREQUENCY, 40 MHz
AMPLITUDE, –10 dBm (50 Input only)
AMPLITUDE, –4.3 dBm (75 Input only)
2. Connect the equipment as shown in Figure 2-46.
Figure 2-46 Second Harmonic Distortion Test Setup
75 Input only: Connect the minimum loss adapter between the LPF
and INPUT 75 .
3. Press Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the controls as follows:
FREQUENCY, Center Freq, 40 MHz
SPAN, 1MHz
AMPLITUDE, –10 dBm (50 Input only)
AMPLITUDE, 44 dBmV (75 Input only)
Attenuation Auto Man 10 dB
BW/Avg, 30 kHz
4. Adjust the synthesized signal generator amplitude to place the peak
of the signal at the reference level.