536 Chapter 3
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
–40 –0.90 dB (112)________ 0.90 dB ±0.081 dB
–44 –0.96 dB (113)________ 0.96 dB ±0.077 dB
–48 –1.02 dB (114)________ 1.02 dB ±0.076 dB
–52 –1.08 dB (115)________ 1.08 dB ±0.080 dB
–56 –1.14 dB (116)________ 1.14 dB ±0.078 dB
–60 –1.20 dB (117)________ 1.20 dB ±0.084 dB
–64 –1.5 dB (118)________ 1.5 dB ±0.081 dB
–68 –1.5 dB (119)________ 1.5 dB ±0.080 dB
–70 –1.5 dB (120)________ 1.5 dB ±0.084 dB
13. Input Attenuation
Switching Uncertainty
Input Attenuation Setting
0 dB –0.3 dB (1)________ 0.3 dB ±0.101 dB
5 dB –0.3 dB (2)________ 0.3 dB ±0.104 dB
15 dB –0.3 dB (3)________ 0.3 dB ±0.102 dB
20 dB –0.3 dB (4)________ 0.3 dB ±0.098 dB
25 dB –0.35 dB (5)________ 0.35 dB ±0.098 dB
30 dB –0.40 dB (6)________ 0.40 dB ±0.096 dB
35 dB –0.45 dB (7)________ 0.45 dB ±0.099 dB
40 dB –0.50 dB (8)________ 0.50 dB ±0.096 dB
45 dB –0.55 dB (9)________ 0.55 dB ±0.099 dB
50 dB –0.60 dB (10)________ 0.60 dB ±0.095 dB
55 dB –0.65 dB (11)________ 0.65 dB ±0.099 dB
60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB
65 dB –0.75 dB (13)________ 0.75 dB ±0.111 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty