224 Chapter 2
Performance Verification Tests
26. Frequency Response (Preamp On): Agilent E4402B
Measuring Frequency Response, Preamp On
1. Refer to Figure 2-39. Remove the reference sensor (Channel A
sensor) from the 20 dB fixed attenuator. Connect the 20 dB fixed
attenuator to the 50 Input of the analyzer using an adapter. Do not
975 MHz
1025 MHz
1075 MHz
1175 MHz
1275 MHz
1375 MHz
1500 MHz
1525 MHz
1675 MHz
1725 MHz
1775 MHz
1825 MHz
1875 MHz
1925 MHz
1975 MHz
2025 MHz
2125 MHz
2325 MHz
2525 MHz
2725 MHz
2925 MHz
2999 MHz
a. This entry is for data taken with the function
generator as source.
b. This entry is for data taken with the synthesized
sweeper as source.
Table 2-60 Source/Splitter Characterization
Frequency
Power Meter Reading Splitter
Tracking
ErrorChannel A Channel B