480 Chapter  3 
Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
Microwave Spectrum 
Analyzer (Option 1DN only) ___________ ___________ ___________
Notes/comments: _____________________________________________________________ 
_____________________________________________________________
Table 3-6  Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty
1. 10 MHz Reference 
Output Accuracy 
  Settability –5.0 Hz (1)________ 5.0 Hz ±293.3 µHz
3. Frequency Readout and 
Marker Frequency 
Count Accuracy
Frequency Readout 
Accuracy
 Center Freq     Span 
   1500 MHz      20 MHz 1499.83 MHz (1)________ 1500.17 MHz ±0 Hz
   1500 MHz      10 MHz 1499.91 MHz (2)________ 1500.09 MHz ±0 Hz
   1500 MHz       1 MHz 1499.991 
MHz
(3)________ 1500.009 
MHz
±0 Hz
   4000 MHz  20 MHz 3999.83 MHz (4)________ 4000.17 MHz ±0 Hz
   4000 MHz  10 MHz 3999.91 MHz (5)________ 4000.09 MHz ±0 Hz
   4000 MHz  1 MHz 3999.991 
MHz
(6)________ 4000.009 
MHz
±0 Hz
Marker Count Accuracy
  Center Freq    Span
    1500 MHz     10 MHz 1499.999999 
MHz
(4)________ 1500.000001 
MHz
±0 Hz
    1500 MHz     1 MHz 1499.999999 
MHz
(5)________ 1500.000001 
MHz
±0 Hz