Chapter 3 433
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
–20 –0.4 dB (26)________ 0.4 dB ±0.078 dB
–24 –0.4 dB (27)________ 0.4 dB ±0.074 dB
–28 –0.4 dB (28)________ 0.4 dB ±0.073 dB
–32 –0.4 dB (29)________ 0.4 dB ±0.077 dB
–36 –0.4 dB (30)________ 0.4 dB ±0.075 dB
–40 –0.4 dB (31)________ 0.4 dB ±0.081 dB
–44 –0.4 dB (32)________ 0.4 dB ±0.077 dB
–48 –0.4 dB (33)________ 0.4 dB ±0.076 dB
–52 –0.4 dB (34)________ 0.4 dB ±0.080 dB
–56 –0.4 dB (35)________ 0.4 dB ±0.078 dB
–60 –0.4 dB (36)________ 0.4 dB ±0.084 dB
–64 –0.4 dB (37)________ 0.4 dB ±0.081 dB
–68 –0.4 dB (38)________ 0.4 dB ±0.080 dB
–72 –0.4 dB (39)________ 0.4 dB ±0.084 dB
–76 –0.4 dB (40)________ 0.4 dB ±0.083 dB
–80 –0.4 dB (41)________ 0.4 dB ±0.089 dB
Cumulative Log Fidelity,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB
–8 –0.38 dB (44)________ 0.38 dB ±0.078 dB
–12 –0.42 dB (45)________ 0.42 dB ±0.075 dB
–16 –0.46 dB (46)________ 0.46 dB ±0.073 dB
–20 –0.50 dB (47)________ 0.50 dB ±0.078 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty