Chapter 2 399
Performance Verification Tests
54. Comms Frequency Response (Option BAC or BAH)
Part 2: Measuring Frequency Response, 800 MHz to 1000 MHz, 
1700 MHz to 2000 MHz (0 dB, 5 dB, 10 dB, 25 dB, and 40 dB 
Attenuation)
Figure 2-75  Comms Frequency Response Test Setup
1. This procedure will be performed several times with various 
combinations of input attenuation, resolution bandwidth, and 
preamp settings as described in Table 2-109. Make 6 copies of Table 
2-110 (7 copies if the analyzer is equipped with the preamp, 
Option 1DS). Label each copy with the information listed in Table 
2-109.
2. Remove the reference sensor (Channel A sensor) from the power 
Table 2-109  Comms Frequency Response Measurement Conditions
Test Measurement Attenuation Res BW Preamp
1
cdmaOne Channel Power 
Accuracy
40 dB 10 kHz Off
2 25 dB 10 kHz Off
3 10 dB 10 kHz Off
4GSM Transmit Power 
Accuracy
40 dB 300 kHz Off
5 5 dB 300 kHz Off
6 cdmaOne Receive Channel 
Power Accuracy
0 dB 10 kHz Off
7 cdmaOne Receive Channel 
Power Accuracy (Preamp On)
0 dB 10 kHz On