430 Chapter 3
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
1490 MHz 10 MHz 1489.91 MHz (2)________ 1490.09 MHz ±0 Hz
1490 MHz 1 MHz 1489.991
MHz
(3)________ 1490.009
MHz
±0 Hz
Marker Count Accuracy
Center Freq Span
1490 MHz 10 MHz 1489.999999
MHz
(4)________ 1490.000001
MHz
±0 Hz
1490 MHz 1 MHz 1489.999999
MHz
(5)________ 1490.000001
MHz
±0 Hz
5. Frequency Span
Accuracy
Span Start Freq
1500 MHz 0 Hz 1185 MHz (1)________ 1215 MHz ±3.06 MHz
100 MHz 10 MHz 79 MHz (2)________ 81 MHz ±204 kHz
100 kHz 10 MHz 79 kHz (3)________ 81 kHz ±204 Hz
100 MHz 800 MHz 79 MHz (4)________ 81 MHz ±204 kHz
100 kHz 800 MHz 79 kHz (5)________ 81 kHz ±204 Hz
100 MHz 1400 MHz 79 MHz (6)________ 81 MHz ±204 kHz
100 kHz 1499 MHz 79 kHz (7)________ 81 kHz ±204 Hz
7. Noise Sidebands
Offset from 1 GHz signal
10 kHz (1)________ –93 dBc/Hz ±2.44 dB
20 kHz (2)________ –100 dBc/Hz ±2.44 dB
30 kHz (3)________ –104 dBc/Hz ±2.44 dB
100 kHz (4)________ –113 dBc/Hz ±2.44 dB
9. System-Related
Sidebands
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty