Chapter 2 371
Performance Verification Tests
48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B (Option 1DN)
1. Complete this step only if more than two hours have elapsed since 
performing a front-panel calibration of the microwave analyzer.
The microwave analyzer should be allowed to warm up for at least 30 
minutes before proceeding.
Complete a front-panel calibration of the microwave analyzer by 
performing the following steps: 
a. Preset the microwave analyzer.
b. Connect a BNC cable between CAL OUTPUT and 50 Ω Input. 
c. Press CAL, REALIGN LO & IF.
d. Set FREQUENCY, 300 MHz.
e. Set SPAN, 20 MHz.
f. Set AMPLITUDE, –10 dBm.
g. Press PEAK SEARCH.
h. Press CAL, REF LVL ADJ and use the ⇑ ⇓ arrows to adjust the DAC 
value to a marker amplitude reading of −10 dBm. Press STORE 
REF LVL.
i. Disconnect the BNC cable from between the CAL OUTPUT and 
50 Ω Input. 
2. Use the Type-N cable to connect the RF Input to the RF OUT of the 
tracking generator as shown in Figure 2-68. Do not connect to the RF 
Input of the microwave analyzer yet.
3. Press Preset on the analyzer under test. Press the Factory Preset 
softkey, if it is displayed. Set the analyzer by pressing the following 
keys:
FREQUENCY, 50 MHz
SPAN, Zero Span
BW/Avg, 30 kHz
System, Alignments, Auto Align, Off
Marker
Source, Tracking Peak (wait for the Peaking message to appear)
Source, Amplitude On, –2 dBm 
Single
4. Set the microwave analyzer by pressing the following keys:
•SPAN, 100 kHz
•AMPLITUDE, 5 dBm 
•AMPLITUDE, Attenuation, 20 dB
•AMPLITUDE, LOG dB/DIV, 10 dB
5. Disconnect the Type-N cable from between the analyzer RF INPUT 
and the tracking generator RF OUT. Refer to Figure 2-68 to connect 
the Type-N cable from the analyzer RF OUT to the microwave