Chapter 3 541
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz
120 kHz 96 kHz (10)________ 144 kHz ±2.17 kHz
9 kHz 7.2 kHz (11)________ 10.8 kHz ±163 Hz
24. Frequency Response Note: Enter data in the appropriate section below depending upon
the ambient temperature at which the test was performed and
installed options.
20 to 30° C
Non-Option UKB
dc Coupled
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.46 dB ±0.19 dB
Minimum Response –0.46 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 0.92 dB ±0.19 dB
Option UKB
dc Coupled
Band 0
(100 Hz to 3.0 GHz)
Maximum Response (1)________ 0.50 dB ±0.19 dB
Minimum Response –0.50 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB ±0.19 dB
dc Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (4)________ 1.5 dB ±0.39 dB
Minimum Response –1.5 dB (5)________ ±0.39 dB
Peak-to-Peak Response (6)________ 2.6 dB ±0.39 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty