Chapter 3 449
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
2nd Harmonic, 20 MHz (2)________ –25 dBc ±2.6 dB
3rd Harmonic, <20 MHz (3)________ –20 dBc ±2.6 dB
3rd Harmonic, 20 MHz (4)________ –25 dBc ±2.6 dB
47. Tracking Generator
Non-Harmonic Spurious
Outputs (Option 1DN or
Option 1DQ only)
Highest Non-Harmonic
Spurious Output
Amplitude
(1)________ –35 dBc ±2.65 dB
50. Gate Delay Accuracy
and Gate Length
Accuracy
(Option 1D6 only)
Minimum Gate Delay 499.9 ns (1)________ 1.5001 µs ±549 ps
Maximum Gate Delay 499.9 ns (2)________ 1.5001 µs ±549 ps
1 µs Gate Length 499.9 ns (3)________ 1.5001 µs ±520 ps
65 ms Gate Length 64.993 ms (4)________ 65.007 ms ±647 ns
51. Gate Mode Additional
Amplitude Error (Option
1D6 only)
Amplitude Error –0.2 dB (1)________ 0.2 dB ±0.023 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty