66 Chapter 2
Performance Verification Tests
7. Noise Sidebands
4. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search)
FREQUENCY, Signal Track (On)
SPAN, 50 kHz
BW/Avg, 100 Hz
Video BW, 30 Hz (Man)
FREQUENCY, Signal Track (Off)
Det/Demod, Detector, Sample
Sweep, Sweep Time, 5 sec
Single
Wait for the completion of a sweep, then press Peak Search (or
Search).
5. Press the following analyzer keys to measure the noise sideband
level at 10 kHz:
Marker, Delta
More, Function, Marker Noise (or Noise)
AMPLITUDE, –10 dBm
FREQUENCY, CF Step, 10 kHz
Center Freq,
SPAN, Zero Span
Single
Record the marker amplitude noise reading in Table 2-10 as the
Upper Sideband Noise Level at 10 kHz.
6. Press the following analyzer keys to measure the noise sideband
level at 10 kHz:
FREQUENCY, Center Freq, ,
Single
Record the marker amplitude noise reading in Table 2-10 as the
Lower Sideband Noise Level at 10 kHz.
7. Press Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1 GHz
SPAN, 10 MHz
AMPLITUDE, Attenuation 10 dB (Man)
8. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search)
FREQUENCY, Signal Track (On)
SPAN, 50 kHz
BW/Avg, 1 kHz
Video BW, 30 Hz (Man)
FREQUENCY, Signal Track (Off)
Det/Demod, Detector, Sample