Chapter 3 501
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
–72 –0.80 dB (18)________ 0.80 dB ±0.084 dB
–76 –0.80 dB (19)________ 0.80 dB ±0.083 dB
–80 –0.80 dB (20)________ 0.80 dB ±0.089 dB
–84 –1.15 dB (21)________ 1.15 dB ±0.086 dB
Incremental Log Fidelity,
Res BW 1 kHz
dB from Ref Level
–4 –0.4 dB (22)________ 0.4 dB ±0.082 dB
–8 –0.4 dB (23)________ 0.4 dB ±0.078 dB
–12 –0.4 dB (24)________ 0.4 dB ±0.075 dB
–16 –0.4 dB (25)________ 0.4 dB ±0.073 dB
–20 –0.4 dB (26)________ 0.4 dB ±0.078 dB
–24 –0.4 dB (27)________ 0.4 dB ±0.074 dB
–28 –0.4 dB (28)________ 0.4 dB ±0.073 dB
–32 –0.4 dB (29)________ 0.4 dB ±0.077 dB
–36 –0.4 dB (30)________ 0.4 dB ±0.075 dB
–40 –0.4 dB (31)________ 0.4 dB ±0.081 dB
–44 –0.4 dB (32)________ 0.4 dB ±0.077 dB
–48 –0.4 dB (33)________ 0.4 dB ±0.076 dB
–52 –0.4 dB (34)________ 0.4 dB ±0.080 dB
–56 –0.4 dB (35)________ 0.4 dB ±0.078 dB
–60 –0.4 dB (36)________ 0.4 dB ±0.084 dB
–64 –0.4 dB (37)________ 0.4 dB ±0.081 dB
–68 –0.4 dB (38)________ 0.4 dB ±0.080 dB
–72 –0.4 dB (39)________ 0.4 dB ±0.084 dB
–76 –0.4 dB (40)________ 0.4 dB ±0.083 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty