530 Chapter  3 
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
       10 MHz (4)________ –137 dBc/Hz ±2.44 dB
9. System-Related 
Sidebands
 Offset from 500 MHz 
signal 
 30 kHz to 230 kHz  (1)________ –65 dBc ±1.30 dB
 –30 kHz to –230 kHz (2)________ –65 dBc ±1.30 dB
10. Residual FM
 1 kHz Res BW
(Non-Option 1D5)
 (1)_________ 150 Hz ±21 Hz
 1 kHz Res BW
(Option 1D5)
 (1)_________ 100 Hz ±21 Hz
 10 Hz Res BW
(Options 1DR 
and 1D5 only)
(2)_________ 2 Hz ±0.27 Hz
11. Sweep Time Accuracy
  Sweep Time
   5 ms  –1.0% (1)________ 1.0% ±0.16%
   20 ms –1.0% (2)________ 1.0% ±0.16%
   100 ms –1.0% (3)________ 1.0% ±0.16%
   1 s –1.0% (4)________ 1.0% ±0.16%
   10 s –1.0% (5)________ 1.0% ±0.16%
   1 ms (Option AYX or B7D 
only)
–1.0% (6)________ 1.0% ±0.16%
   500 µs (Option AYX or 
B7D only)
–1.0% (7)________ 1.0% ±0.16%
   100 µs (Option AYX or 
B7D only)
–1.0% (8)________ 1.0% ±0.16%
12. Display Scale Fidelity
Table 3-10  Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty