454 Chapter 3
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
1 kHz Res BW,
(Non-Option 1D5)
(1)_________ 150 Hz ±21 Hz
1 kHz Res BW,
(Option 1D5)
(1)_________ 100 Hz ±21 Hz
10 Hz Res BW
(Options 1DR and
1D5 only)
(3)_________ 2 Hz ±0.27 Hz
11. Sweep Time Accuracy
Sweep Time
5 ms –1.0% (1)________ 1.0% ±0.16%
20 ms –1.0% (2)________ 1.0% ±0.16%
100 ms –1.0% (3)________ 1.0% ±0.16%
1 s –1.0% (4)________ 1.0% ±0.16%
10 s –1.0% (5)________ 1.0% ±0.16%
1 ms (Option AYX or B7D
only)
–1.0% (6)________ 1.0% ±0.16%
500 µs (Option AYX or
B7D only)
–1.0% (7)________ 1.0% ±0.16%
100 µs (Option AYX or
B7D only)
–1.0% (8)________ 1.0% ±0.16%
12. Display Scale Fidelity
Cumulative Log Fidelity,
Res BW 1 kHz
dB from Ref Level
–4 –0.30 dB (1)________ 0.30 dB ±0.082 dB
–8 –0.30 dB (2)________ 0.30 dB ±0.078 dB
–12 –0.40 dB (3)________ 0.40 dB ±0.075 dB
–16 –0.40 dB (4)________ 0.40 dB ±0.073 dB
–20 –0.40 dB (5)________ 0.40 dB ±0.078 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty