454 Chapter  3 
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
  1 kHz Res BW,
(Non-Option 1D5)
 (1)_________ 150 Hz ±21 Hz
  1 kHz Res BW,
(Option 1D5)
 (1)_________ 100 Hz ±21 Hz
  10 Hz Res BW
(Options 1DR and
1D5 only)
(3)_________ 2 Hz ±0.27 Hz
11. Sweep Time Accuracy
  Sweep Time
   5 ms  –1.0% (1)________ 1.0% ±0.16%
   20 ms –1.0% (2)________ 1.0% ±0.16%
   100 ms –1.0% (3)________ 1.0% ±0.16%
   1 s –1.0% (4)________ 1.0% ±0.16%
   10 s –1.0% (5)________ 1.0% ±0.16%
   1 ms (Option AYX or B7D 
only)
–1.0% (6)________ 1.0% ±0.16%
   500 µs (Option AYX or 
B7D only)
–1.0% (7)________ 1.0% ±0.16%
   100  µs (Option AYX or 
B7D only)
–1.0% (8)________ 1.0% ±0.16%
12. Display Scale Fidelity
Cumulative Log Fidelity,
Res BW ≥1 kHz
dB from Ref Level
–4 –0.30 dB (1)________ 0.30 dB ±0.082 dB
–8  –0.30 dB (2)________ 0.30 dB ±0.078 dB
–12  –0.40 dB (3)________ 0.40 dB ±0.075 dB
–16  –0.40 dB (4)________ 0.40 dB ±0.073 dB
–20  –0.40 dB (5)________ 0.40 dB ±0.078 dB
Table 3-4  Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty