Chapter 2 253
Performance Verification Tests
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B

30.On the analyzer, press AMPLITUDE, Presel Center.

31.On the analyzer, press the following:

Marker, More, Marker All Off

32.Repeat step 5 through step 11 for the synthesized sweeper CW

frequencies listed in Table 2-68 for Band 4 for the 21 GHz analyzer

center frequency.

Table 2-68 Other Input-Related Spurious Responses

Band
Analyzer
Center
Frequency
(GHz)
Synthesized
Sweeper CW
Frequency
(MHz)
Mixer
Level
(dBm)
Test Record
Entry
Mkr1 or
Mkr2
Amplitude
02.0 2042.8a–20 1)
2.0 2642.8a–20 2)
2.0 1820.8b–20 3)
2.0 278.5b–20 4)
Note: The following data applies only to the Agilent E4404B, E4405B,
E4407B and E4408B
02.0 5600.0c–10 5)
2.0 6242.8c–10 6)
14.0 4042.8a–20 7)
4.0 4642.8a–20 8)
4.0 3742.9b–20 9)
4.0 2242.8c–10 10)
Note: The following data applies only to the Agilent E4405B, E4407B and
E4408B
29.0 9042.8a–20 11)
9.0 9642.8a–20 12)
9.0 4982.1b–20 13)
9.0 9342.8c–10 14)
Note: The following data applies only to the Agilent E4407B and E4408B
315.0 15042.8a–20 15)