
Chapter 2 103
Performance Verification Tests
13. Input Attenuation Switching Uncertainty
4. For each attenuation error value in Column 4 of Table 2-19, calculate 
the attenuation errors by subtracting the difference between the 
Table 2-18 Total Actual Attenuation and Total Nominal Attenuation 
from the difference between the Total Actual Attenuation and Total 
Nominal Attenuation at 55 dB. Note that the total nominal 
attenuations listed in Table 2-19 are in a different order than those 
listed in Table 2-18.
AttenErr = (ActAtten(55 dB) - 55 dB) - (ActAtten(X dB) - 
NomAtten(X dB))
Where:
Example of attenuation error calculation for 35 dB nominal 
attenuation:
Setup for Switching Uncertainty Measurement
5. Connect the equipment as indicated in Figure 2-13. The 6 dB fixed 
attenuator (or minimum loss pad for 75 Ω input analyzers) should be 
connected directly to the input connector of the analyzer.
6. Press Preset on the analyzer. Press the Factory Preset softkey, if it is 
displayed. Press System, Alignments, Auto Align, Off. Set the controls 
as follows:
FREQUENCY, 50 MHz
SPAN, 100 kHz 
AMPLITUDE, –55 dBm (50 Ω Input only)
AMPLITUDE, –6.2 dBmV (75 Ω Input only)
AMPLITUDE, Attenuation, 10 dB 
AMPLITUDE, Scale/Div, 2 dB 
BW/Avg, 30 kHz
BW/Avg, Video BW, 100 Hz