Chapter  3 631
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
 50 Ω, 100 Hz RBW
       400 kHz (11)________ –123 dBm ±1.23 dB
       1 MHz to 10 MHz (12)________ –123 dBm ±1.23 dB
       10 MHz to 500 MHz (13)________ –127 dBm ±1.23 dB
       500 MHz to 1 GHz (14)________ –125 dBm ±1.23 dB
       1 GHz to 1.5 GHz (15)________ –121 dBm ±1.23 dB
75 Ω, 1 kHz RBW
       1 MHz to 10 MHz (21)________ –63 dBmV ±1.23 dB
       10 MHz to 500 MHz (22)________ –65 dBmV ±1.23 dB
       500 MHz to 1 GHz (23)________ –60 dBmV ±1.23 dB
       1 GHz to 1.5 GHz (24)________ –53 dBmV ±1.23 dB
75 Ω, 100 Hz RBW
        1 MHz to 10 MHz (29)________ –72 dBmV ±1.23 dB
        10 MHz to 500 MHz (30)________ –74 dBmV ±1.23 dB
        500 MHz to 1 GHz (31)________ –68 dBmV ±1.23 dB
        1 GHz to 1.5 GHz (32)________ –61 dBmV ±1.23 dB
39. Residual Responses Note: Enter data in the appropriate section below depending upon 
the input impedance and serial number of the analyzer.
50 Ω, 150 kHz to 1.5 GHz (1)________ –90 dBm ±0.87 dB
75 Ω, 1 MHz to 1.5 GHz   (1)________ –36 dBmV ±0.87 dB
41. Tracking Generator 
Absolute Amplitude 
and Vernier Accuracy
Note: Enter data in the appropriate section below depending upon 
the input impedance and serial number of the analyzer.
50 Ω (Option 1DN)
Absolute Amplitude 
Accuracy
–0.5 dB (1)________  0.5 dB ±0.096 dB
Positive Vernier Accuracy (2)________  0.75 dB ±0.008 dB
Table 3-16  Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty