Chapter 3 463
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
3 Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (16)________ 0.3 dB ±0.048 dB
1Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (17)________ 0.3 dB ±0.048 dB
18. Absolute Amplitude
Accuracy (Reference
Settings)
Log, Preamp Off –0.34 dB (1)________ 0.34 dB ±0.09 dB
Lin, Preamp Off –0.34 dB (2)________ 0.34 dB ±0.09 dB
Log, Preamp On
(Option 1DS only)
–0.37 dB (3)________ 0.37 dB ±0.09 dB
Lin, Preamp On
(Option 1DS only)
–0.37 dB (4)________ 0.37 dB ±0.09 dB
20. Overall Absolute
Amplitude Accuracy
0 dBm Reference Level
0 dBm input –0.54 dB (1)________ 0.54 dB ±0.234 dB
–10 dBm input –0.54 dB (2)________ 0.54 dB ±0.115 dB
–20 dBm input –0.54 dB (3)________ 0.54 dB ±0.110 dB
–30 dBm input –0.54 dB (4)________ 0.54 dB ±0.109 dB
–40 dBm input –0.54 dB (5)________ 0.54 dB ±0.109 dB
–50 dBm input –0.54 dB (6)________ 0.54 dB ±0.109 dB
20 dBm Reference Level
–20 dBm input –0.54 dB (7)________ 0.54 dB ±0.109 dB
–30 dBm input –0.54 dB (8)________ 0.54 dB ±0.109 dB
–40 dBm input –0.54 dB (9)________ 0.54 dB ±0.109 dB
–50 dBm input –0.54 dB (10)________ 0.54 dB ±0.109 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty