Chapter 2 37
Performance Verification Tests
Tests included in this section:
Calibration
To perform calibration:
1. Run all performance verification tests listed in Column 1 of Table 2-1 
that are applicable to the analyzer being calibrated and the installed 
options. Refer to the Performance Verification test tables in 
Chapter 1 for a list of the tests that are applicable to the analyzer 
being calibrated.
2. If any of the performance verification tests fail, perform the 
appropriate calibration adjustments listed in Column 2 of Table 2-1 
which corresponds to the failure.
3. Repeat all of the performance verification tests listed in Column 1 of 
Table 2-1 if any calibration adjustments were made in step 2. This 
will confirm that there is no interaction between adjustments which 
could negatively impact analyzer performance.
Table 2-1 lists the performance verification tests and adjustments 
needed for calibration. The performance tests are located in the 
following pages of this chapter. Adjustment information is located in the 
service guide.