630 Chapter  3 
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
   Minimum Response –1.0 dB (2)________ ±0.11 dB
   Peak-to-Peak Response (3)________ 2.0 dB ±0.11 dB
28. Other Input-Related 
Spurious Responses
   Input Frequency
        542.8 MHz (1)________ –65 dBc ±.889 dB
        510.7 MHz (2)________ –65 dBc  ±.889 dB
        1310.7 MHz (3)________ –45 dBc ±.889 dB
30. Spurious Responses  Note: Enter the results in the appropriate lines below based upon the 
input impedance of the analyzer. Test Record Entry 2 does not apply 
to Agilent E4411B.
     50 MHz TOI, 50 Ω 7.5 dBm (1)________   ±0.93 dB
     50 MHz TOI,  75 Ω 56.25 dBmV (1)________ ±0.93 dB
     40 MHz SHI, 50 Ω 35 dBm (3)________ ±.050 dB
     40 MHz SHI, 75 Ω 83.75 dBmV (3)________ ±.050 dB
33. Gain Compression Note: Test Record Entry 2 does not apply to Agilent E4411B.
Test Frequency
  53 MHz (1)________  1.0 dB ±0.123 dB
  1403 MHz (3)________  1.0 dB ±0.123 dB
35. Displayed Average 
Noise Level
Note: Enter data in the appropriate section below depending upon 
the input impedance and serial number of the analyzer.
50 Ω, 1 kHz RBW
       400 kHz (1)________ –115 dBm ±1.23 dB
       1 MHz to 10 MHz (2)________ –115 dBm ±1.23 dB
       10 MHz to 500 MHz (3)________ –119 dBm ±1.23 dB
       500 MHz to 1 GHz (4)________ –117 dBm ±1.23 dB
       1 GHz to 1.5 GHz (5)________ –113 dBm ±1.23 dB
Table 3-16  Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty