Chapter 2 249
Performance Verification Tests
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B
29. Other Input-Related Spurious Responses:
Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B
This test measures the ability of the analyzer to reject image, multiple,
and out-of-band responses. A synthesized source and the analyzer are
set to the same frequency and the amplitude of the source is set to –10
dBm and –20 dBm. A marker amplitude reference is set on the analyzer
for each source amplitude setting. The source is then tuned to several
different frequencies which should generate image, multiple, and
out-of-band responses. At each source frequency, the source amplitude
is set to the appropriate amplitude and the amplitude of the response, if
any, is measured using the analyzer marker functions.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Power meter
RF power sensor, (Agilent E4402B, E4403B)
Microwave power sensor (Agilent E4404B, E4405B, E4407B,
E4408B)
RF power splitter (Agilent E4402B, E4403B)
Microwave power splitter (Agilent E4404B, E4405B, E4407B,
E4408B)
Adapter, Type-N (m) to Type-N (m)
Adapter, Type-N (m) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in)
Procedure
Band 0
1. Zero and calibrate the power meter and power sensor in log mode
(power reads out in dBm), as described in the power meter operation
manual. Enter the power sensor 2 GHz calibration factor into the
power meter.
2. Preset the synthesized sweeper and set the controls as follows:
CW, 2000 MHz
POWER LEVEL, –4 dBm
3. Connect the equipment as shown in Figure 2-44. The analyzer
provides the 10 MHz reference for the synthesized sweeper.