482 Chapter  3 
Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
   20 ms –1.0% (2)________ 1.0% ±0.16%
   100 ms –1.0% (3)________ 1.0% ±0.16%
   1 s –1.0% (4)________ 1.0% ±0.16%
   10 s –1.0% (5)________ 1.0% ±0.16%
12. Display Scale Fidelity
Cumulative Log Fidelity,
Res BW ≥1 kHz
dB from Ref Level
–4  –0.34 dB (1)________ 0.34 dB ±0.082 dB
–8  –0.38 dB (2)________ 0.38 dB ±0.078 dB
–12  –0.42 dB (3)________ 0.42 dB ±0.075 dB
–16  –0.46 dB (4)________ 0.46 dB ±0.073 dB
–20  –0.50 dB (5)________ 0.50 dB ±0.078 dB
–24  –0.54 dB (6)________ 0.54 dB ±0.074 dB
–28  –0.58 dB (7)________ 0.58 dB ±0.073 dB
–32  –0.62 dB (8)________ 0.62 dB ±0.077 dB
–36  –0.66 dB (9)________ 0.66 dB ±0.075 dB
–40  –0.70 dB (10)________ 0.70 dB ±0.081 dB
–44  –0.74 dB (11)________ 0.74 dB ±0.077 dB
–48  –0.78 dB (12)________ 0.78 dB ±0.076 dB
–52  –0.82 dB (13)________ 0.82 dB ±0.080 dB
–56  –0.86 dB (14)________ 0.86 dB ±0.078 dB
–60  –0.90 dB (15)________ 0.90 dB ±0.084 dB
–64  –0.94 dB (16)________ 0.94 dB ±0.081 dB
–68  –0.98 dB (17)________ 0.98 dB ±0.080 dB
–72  –1.02 dB (18)________ 1.02 dB ±0.084 dB
Table 3-6  Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty