360 Chapter 2 
Performance Verification Tests
46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, 
E4405B, E4407B, and E4408B (Option 1DN)
46. Tracking Generator Harmonic Spurious
Outputs: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B (Option 1DN)
The measurement for tracking generator harmonic spurious outputs 
determines the maximum level of tracking generator harmonics. The 
tracking generator output is connected to the input of a microwave 
analyzer, then tuned to several different frequencies as the amplitude of 
the second and third harmonics (relative to the fundamental) are 
measured at each frequency.
There are no related adjustment procedures for this performance test. 
Equipment Required 
Microwave analyzer 
Cable, Type-N, 62 cm (24 in)
Cable, BNC to BNC, 23 cm (9 in)
Adapter, Type-N (m) to BNC (f)
Procedure 
NOTE The following steps are for an Agilent 8563E microwave analyzer, the 
steps may be different if you are using another microwave analyzer. 
1. Use the Type-N cable to connect the RF INPUT to the tracking 
generator RF OUT as shown in Figure 2-66. Do not connect the 
Type-N cable to the microwave analyzer yet. Connect the 10 MHz 
Reference from the output of the microwave analyzer to the 10 MHz 
Reference Input of the analyzer being tested.