580 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Option UKB
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (22)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (23)________ ±0.11 dB
Peak-to-Peak Response (24)________ 2.0 dB ±0.11 dB
Option UKB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (25)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (26)________ ±0.11 dB
Peak-to-Peak Response (27)________ 2.0 dB ±0.11 dB
ac Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (28)________ 2.5 dB ±0.39 dB
Minimum Response –2.5 dB (29)________ ±0.39 dB
Peak-to-Peak Response (30)________ 3.0 dB ±0.39 dB
ac Coupled
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (31)________ 3.0 dB ±0.40 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty