
294 Chapter 2 
Performance Verification Tests
34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B
34. Gain Compression: Agilent E4404B,
E4405B, E4407B, and E4408B
This test verifies the ability of the analyzer to measure relatively 
low-amplitude signals in the presence of higher-amplitude signals. Gain 
compression is measured by applying two signals, separated by a 
defined amount in frequency. The higher-amplitude signal is set to yield 
the specified total power at the input mixer (the power at the input 
mixer is defined as the input power level minus the input attenuation). 
The lower-amplitude signal is set at least 35 dB below the 
higher-amplitude signal, such that its power does not significantly add 
to the total power. The higher-amplitude signal is turned off and the 
lower-amplitude signal level is measured. This is the uncompressed 
amplitude. 
The higher-amplitude signal is turned on and the amplitude of the 
lower-amplitude signal is again measured. This is the compressed 
amplitude. The difference between the uncompressed and compressed 
amplitude is the measured gain compression.
There are no related adjustment procedures for this performance test. 
Equipment Required 
Synthesized sweeper (2 required)
Power meter, dual channel
Microwave power sensor 
Directional bridge 
Directional coupler 
Cable, BNC, 120-cm (48-in) 
Cable, APC 3.5 (m) (2 required)
Adapter, Type-N (m) to Type-N (m) 
Adapter, Type-N (m) to APC 3.5 (f) (3 required)
Adapter, Type-N (m) to SMA (m)
Additional Equipment for Option BAB
Adapter, Type-N (m), to APC 3.5 (f)
Procedure 
1. Zero and calibrate the power meter and power sensor combination in 
log mode (power reads out in dBm) as described in the power meter 
operation manual.
2. Connect the equipment as shown in Figure 2-50, with the load port 
of the directional bridge connected to the power sensor. The 
directional bridge should be used for measurements of frequencies