628 Chapter  3 
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
     –30 dBm input –0.6 dB (4)________  0.6 dB ±0.109 dB
     –40 dBm input –0.6 dB (5)________  0.6 dB ±0.109 dB
     –50 dBm input –0.6 dB (6)________  0.6 dB ±0.109 dB
–20 dBm Reference Level
     –20 dBm input –0.6 dB (7)________  0.6 dB ±0.109 dB
     –30 dBm input –0.6 dB (8)________  0.6 dB ±0.109 dB
     –40 dBm input –0.6 dB (9)________  0.6 dB ±0.109 dB
     –50 dBm input –0.6 dB (10)________  0.6 dB ±0.109 dB
–40 dBm Reference Level
     –40 dBm input –0.6 dB (11)________  0.6 dB ±0.087 dB
     –50 dBm input –0.6 dB (12)________  0.6 dB ±0.087 dB
–50 dBm Reference Level
     –50 dBm input –0.6 dB (13)________  0.6 dB ±0.087 dB
21. Resolution Bandwidth 
Accuracy
Resolution Bandwidth
 5 MHz 3.5 MHz (1)________ 6.5 MHz ±110 kHz
 3 MHz 2.55 MHz (2)________ 3.45 MHz ±66 kHz
 1 MHz 0.85 MHz (3)________ 1.15 MHz ±22 kHz
300 kHz 255 kHz (4)________ 345 kHz ±6.6 kHz
100 kHz 85 kHz (5)________ 115 kHz ±2.2 kHz
30 kHz 25.5 kHz (6)________ 34.5 kHz ±660 Hz
10 kHz 8.5 kHz (7)________ 11.5 kHz ±220 Hz
3 kHz 2.55 kHz (8)________ 3.45 kHz ±66 Hz
1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz
120 kHz 96 kHz (10)________ 144 kHz ±2.17 kHz
Table 3-16  Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty