Chapter 2 353
Performance Verification Tests
44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B (Option 1DN)
12.See Figure 2-63. Disconnect the power sensor from the RF Out 50
on the analyzer.

Tracking Generator Level Flatness, Center

Frequency 100 kHz

1. Refer to Figure 2-64 to set up the equipment.
Figure 2-64 Tracking Generator Level Flatness Test Setup, 100 kHz
Set up the digital multimeter as follows.
1900 MHz 2000
2050 MHz 2000
2200 MHz 2000
2350 MHz 2000
2500 MHz 3000
2650 MHz 3000
2800 MHz 3000
2950 MHz 3000
3000 MHz 3000
Table 2-97 Tracking Generator Level Flatness Worksheet, 100 kHz
Center
Frequency
Level Flatness
(dB)
Cal Factor
(MHz)
Parameter Setting
AC/DC AC Volts
Impedance & Units: