556 Chapter 3
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
−40 dBm –1.88 dB (35)________ 1.88 dB ±0.161 dB
−60 dBm –1.88 dB (36)________ 1.88 dB ±0.161 dB
−70 dBm –1.88 dB (37)________ 1.88 dB ±0.161 dB
−80 dBm –1.88 dB (38)________ 1.88 dB ±0.161 dB
−90 dBm –2.95 dB (39)________ 2.95 dB ±0.161 dB
−100 dBm –2.95 dB (40)________ 2.95 dB ±0.161 dB
PCS Band
Input Amplitude
−40 dBm –1.88 dB (41)________ 1.88 dB ±0.161 dB
−60 dBm –1.88 dB (42)________ 1.88 dB ±0.161 dB
−70 dBm –1.88 dB (43)________ 1.88 dB ±0.161 dB
−80 dBm –1.88 dB (44)________ 1.88 dB ±0.161 dB
−90 dBm –2.95 dB (45)________ 2.95 dB ±0.161 dB
−100 dBm –2.95 dB (46)________ 2.95 dB ±0.161 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty