Chapter 2 101
Performance Verification Tests
13. Input Attenuation Switching Uncertainty
Figure 2-13 Input Attenuator Switching Uncertainty Test Setup
CAUTION Use only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
connectors, or the connectors will be damaged.
Procedure
Calculate Actual Attenuation Values
1. From the calibration data supplied with the 1 dB step attenuator,
enter into Column 2 of Table 2-18 the actual attenuation for the
corresponding nominal attenuation settings. If the calibration data
does not indicate an actual attenuation value for the 0 dB setting,
enter 0 dB.
NOTE The Agilent 8494G programmable attenuator has four attenuator
sections consisting of 1 dB, 2 dB, 4 dB, and 4 dB attenuators. If using
the Agilent 8494G programmable attenuator, enter the calibration data
for the section three 4 dB step rather than the section four 4 dB step.
2. From the calibration data supplied with the 10 dB step attenuator,
enter into Column 4 of Table 2-18 the actual attenuation for the
corresponding nominal attenuation settings. If the calibration data
does not indicate an actual attenuation value for the 0 dB setting,
enter 0 dB.
NOTE The Agilent 8496G programmable attenuator has four attenuator
sections consisting of 10 dB, 20 dB, 40 dB, and 40 dB attenuators. If