Chapter 2 105
Performance Verification Tests
13. Input Attenuation Switching Uncertainty
Measure Switching Uncertainty
Perform step 12 to step 15 for each measurement value in Table 2-19.
12.Set the 1 dB and 10 dB step attenuators to the Total Nominal 
Attenuation setting value as indicated in Column 3 of Table 2-19 for 
the various analyzer attenuation settings for each measurement. 
Table 2-18 may be used as a reference for setting the step 
attenuators to achieve the desired total nominal attenuation.
13.Similarly, set the corresponding analyzer attenuation and reference 
level settings as indicated in Column 1 and Column 2 of Table 2-19.
14.Press Single, then Peak Search (or Search) and record the marker 
amplitude reading as the Marker Delta Reading in Column 6 of 
Table 2-19.
15.Calculate the Switching Error (Table 2-19, Column 7) by subtracting 
the Ideal Marker Delta Reading and the Attenuation Error from the 
Marker Delta Reading. Record the result in the performance 
verification test record.
35 dB –30 dBm / 
18.8 dBmV
30 dB 25 dB 7) 
40 dB –25 dBm / 
23.8 dBmV
25 dB 30 dB 8) 
45 dB –20 dBm / 
28.8 dBmV
20 dB 35 dB 9) 
50 dB –15 dBm / 
33.8 dBmV
15 dB 40 dB 10) 
55 dB –10 dBm / 
38.8 dBmV
10 dB 45 dB 11) 
60 dB –5 dBm / 43.8 
dBmV
5 dB 50 dB 12) 
65 dBa0 dBm /
48.8 dBmV
0 dB 55 dB 13) 
a. Does not apply to Agilent E4401B or E4411B.
Table 2-19  Input Attenuation Switching Uncertainty Worksheet
Column 1 Column 2 Column 3 Column 4 Column 5 Column 6 Column 7
Analyzer 
Internal 
Attenu-
ation 
Setting
Analyzer 
Reference 
Level 
Setting
50 W Input/ 
75 W Input
Total 
Nominal 
Attenu-
ation
Setting
Attenu-
ation 
Error 
(dB)
Ideal 
Marker 
Delta 
Reading
Marker 
Delta 
Reading
(dB)
Test 
Record 
Entry – 
Switching 
Error
(dB)