Chapter  3 615
Performance Verification Test Records
Agilent E4408B Performance Verification Test Record
–6 dB –0.2 dB (13)________ 0.2 dB ±0.008 dB
–7 dB –0.2 dB (14)________ 0.2 dB ±0.008 dB
–8 dB –0.2 dB (15)________ 0.2 dB ±0.008 dB
–9 dB –0.2 dB (16)________ 0.2 dB ±0.008 dB
–10 dB –0.2 dB (17)________ 0.2 dB ±0.008 dB
44. Tracking Generator 
Level Flatness (Option 
1DN only)
Positive Level Flatness, 
<1 MHz
(1)________  3.0 dB ±0.284 dB
 Negative Level Flatness, 
<1 MHz
–3.0 dB (2)________ ±0.284 dB
 Positive Level Flatness, 
1 MHz to 10 MHz
(3)________  3.0 dB ±0.142 dB
 Negative Level Flatness, 
1 MHz to 10 MHz
–3.0 dB (4)________ ±0.142 dB
 Positive Level Flatness, 
>10 MHz to 1.5 GHz
(5)________  2.0 dB ±0.142 dB
 Negative Level Flatness, 
>10 MHz to 1.5 GHz
–2.0 dB (6)________ ±0.142 dB
 Positive Level Flatness, 
>1.5 GHz
(7)________  2.0 dB ±0.142 dB
 Negative Level Flatness, 
>1.5 GHz
–2.0 dB (8)________ ±0.142 dB
46. Tracking Generator 
Harmonic Spurious 
Outputs 
(Option 1DN only)
2nd Harmonic, <20 kHz (1)________ –15 dBc ±2.6 dB
2nd Harmonic, ≥20 kHz (2)________ –25 dBc ±2.6 dB
Table 3-14  Agilent E4408B Performance Verification Test Record
Agilent Technologies
Model E4408B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty