24 Chapter  1  
Calibrating  
Before You Start
Before You Start
This brief procedure should be performed before starting the 
performance verification tests:
1. Switch the analyzer on and let it warm up for five minutes.
2. If the analyzer is an Agilent E4402B, E4403B, E4404B, E4405B, 
E4407B, or E4408B, connect a cable from AMPTD REF OUT to the 
50 Ω Input.
3. Press System, Alignments, Align Now, All, and wait for the auto 
alignments to finish.
4. Read the remainder of this section before you start any of the tests, 
and make a copy of the Performance Verification Test Record 
described below in "Recording the test results."
Recording the test results
Performance verification test records, for each analyzer, are provided in 
the chapter following the tests.
Each test result is identified as a TR Entry in the performance tests 
and on the performance verification test record. We recommend that 
you make a copy of the performance verification test record, record the 
test results on the copy, and keep the copy for your calibration test 
record. This record could prove valuable in tracking gradual changes in 
test results over long periods of time.
Performing Self-Alignment
Perform a complete self-alignment at least once per day, or if the 
analyzer fails a verification test. To perform a self-alignment, press 
System, Alignment, Align Now, All. The instrument must be up to 
operating temperature in order for this test to be valid. If the analyzer 
continuously fails one or more specifications, complete any remaining 
tests and record all test results on a copy of the test record. Then refer 
to Chapter 4 , “If You Have a Problem,” for instructions on how to solve 
the problem.