182 Chapter 2 
Performance Verification Tests
23. Frequency Response, Agilent E4402B and E4403B
Measuring Frequency Response Less Than or Equal to 100 kHz 
1. Connect the equipment as shown in Figure 2-31.
Figure 2-31  Frequency Response Test Setup, ≤100 kHz
2. Set the function generator controls as follows: 
FREQUENCY, 100 kHz
AMPLITUDE, –4 dBm
Amptd Increment, 0.05 dB
3. Set the DVM as follows:
4. On the analyzer, press FREQUENCY, 100 kHz.
5. If the analyzer has Option UKB, press Input/Output (or Input), 
Coupling (DC). Then press SPAN, 100 Hz, BW/Avg, Res BW, 10 Hz, Video 
BW, 10 Hz.
6. Adjust the function generator amplitude until the DVM reading is 
–10 dBm ±0.1 dB.
7. Record the actual DVM reading in Table 2-46 as the DVM Amplitude 
reading.
8. On the analyzer, press Peak Search (or Search), Marker, Delta. 
9. Set the analyzer center frequency to the next frequency listed in 
Table 2-46. 
10.Set the function generator frequency to the next frequency listed in 
Function Synchronous AC Volts
Math dBm
RES Register 50 Ω
Front/Rear Terminals Front
Range Auto