64 Chapter 2
Performance Verification Tests
7. Noise Sidebands
Procedure
This performance test consists of five parts:
Part 1: Noise Sideband Suppression at 10 kHz (measured in 1 kHz
RBW)
Part 2: Noise Sideband Suppression at 10 kHz (measured in 100 Hz
RBW)
Part 3: Noise Sideband Suppression at 20 kHz
Part 4: Noise Sideband Suppression at 30 kHz
Part 5: Noise Sideband Suppression at 100 kHz (Agilent E4401B and
E4411B only)
For E4401B or E4411B, do Parts 1, 3, 4 and 5.
For E4402B, E4404B, E4405B or E4407B with Option 1DR and
serial prefixes greater than or equal to US4510, SG4510, or MY4510,
do Parts 2, 3 and 4.
For all other ESA series analyzers do Parts 1, 3 and 4.
A worksheet is provided at the end of this procedure for calculating the
noise sideband suppression.
Part 1: Noise Sideband Suppression at 10 kHz (measured in
1 kHz RBW)
1. Perform the following steps to set up the equipment:
Set the synthesized signal generator controls as follows:
FREQUENCY, 1000 MHz
AMPLITUDE, 0dBm (50 Input only)
AMPLITUDE, 6 dBm (75 Input only)
AM OFF
FM OFF
2. Connect the equipment as shown in Figure 2-7.
3. Press Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1 GHz
SPAN, 10 MHz
AMPLITUDE, Attenuation 10 dB (Man)
4. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search)
FREQUENCY, Signal Track (On)
SPAN, 50 kHz
BW/Avg, 1kHz
Video BW, 30 Hz (Man)
FREQUENCY, Signal Track (Off)