Chapter  3 587
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
 14003 MHz (7)________  1.0 dB ±0.123 dB
38. Displayed Average 
Noise Level
Note: Enter results with preamp on in the appropriate section based 
upon the ambient temperature when the test was performed. 
1 kHz RBW, Preamp Off
          10 MHz to 1 GHz (1)________ –116 dBm ±1.23 dB
          1 GHz to 2 GHz (2)________ –116 dBm ±1.23 dB
          2 GHz to 3 GHz (3)________ –112 dBm ±1.23 dB
          3 GHz to 6 GHz (4)________ –112 dBm ±1.23 dB
          6 GHz to 12 GHz (5)________ –111 dBm ±1.23 dB
        12 GHz to 22 GHz (6)________ –107 dBm ±1.23 dB
       22 GHz to 26.5 GHz (7)________ –106 dBm ±1.23 dB
1kHz RBW, Preamp On,
0 to 55° C
          10 MHz to 1 GHz (8)________ –131 dBm ±1.23 dB
          1 GHz to 2 GHz (9)________ –131 dBm ±1.23 dB
          2 GHz to 3 GHz (10)________ –133 dBm ±1.23 dB
10 Hz RBW, Preamp Off
          10 MHz to 1 GHz (11)________ –135 dBm ±1.23 dB
          1 GHz to 2 GHz (12)________ –135 dBm ±1.23 dB
          2 GHz to 3 GHz (13)________ –131 dBm ±1.23 dB
          3 GHz to 6 GHz (14)________ –131 dBm ±1.23 dB
          6 GHz to 12 GHz (15)________ –130 dBm ±1.23 dB
       12 GHz to 22 GHz (16)________ –126 dBm ±1.23 dB
      22 GHz to 26.5 GHz (17)________ –125 dBm ±1.23 dB
10 Hz RBW, Preamp On,
0 to 55° C
Table 3-12  Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty