Chapter 2 319
Performance Verification Tests
37. Displayed Average Noise Level: Agilent E4404B and E4405B
17.If the analyzer has Option 1DN, 50 Ω tracking generator, do the 
following:
a. On the analyzer, press BW/Avg, Res BW, 1 kHz.
b. Press Source, Amplitude, 0 dBm.
c. Connect a 50 Ω termination to the RF OUT 50 Ω.
Measurement Sequence
The following model-specific DANL Measurement Sequence tables list 
the procedures to be performed and the parameters to be used in each 
procedure. Also listed in the tables are test record entry numbers for 
recording the results in the performance verification test record.
1. Perform all of the following steps (through step 7) that apply to your 
analyzer using the appropriate subsets in Table 2-87 or Table 2-88. 
Then record the display line amplitude setting as the indicated Test 
Record entry in the performance verification test record.
2. If the minimum RBW of the analyzer is 1 kHz, perform those 
procedures listed as Subset A in Table 2-87 or Table 2-88.
3. If the minimum RBW of the analyzer is 1 kHz and Option 1DS (RF 
Preamplifier) is installed, also perform those procedures listed in 
Subset B in Table 2-87 or Table 2-88.
4. If the minimum RBW of the analyzer is 10 Hz and Option 1DN (3.0 
GHz Tracking Generator) is installed, perform those procedures 
listed in Subset A in Table 2-87 or Table 2-88.
5. If the minimum RBW of the analyzer is 10 Hz and both Option 1DS 
(RF Preamplifier) and Option 1DN (3.0 GHz Tracking Generator) 
are installed, also perform those procedures listed in Subset B in 
Table 2-87 or Table 2-88.
6. If the minimum RBW of the analyzer is 10 Hz, also perform those 
procedures listed in Subset C in Table 2-87 or Table 2-88.
7. If the minimum RBW of the analyzer is 10 Hz and Option 1DS (RF 
Preamplifier) is installed, also perform those procedures listed in 
Subset D in Table 2-87 or Table 2-88.
8. After performing all applicable DANL measurement procedures, 
continue with Remove Reference Level Offset.
Table 2-87  DANL Measurement Sequence, E4404B
Subset Procedure
Procedure Parameters Test 
Record 
Entrya
Start 
Freq
Stop 
Freq
Test 
RBW
Preamp 
State
A Measure DANL 10 MHz 1 GHz 1 kHz Off 1)