622 Chapter 3
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
–76 –0.4 dB (40)________ 0.4 dB ±0.083 dB
–80 –0.4 dB (41)________ 0.4 dB ±0.089 dB
Cumulative Log Fidelity,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB
–8 –0.38 dB (44)________ 0.38 dB ±0.078 dB
–12 –0.42 dB (45)________ 0.42 dB ±0.075 dB
–16 –0.46 dB (46)________ 0.46 dB ±0.073 dB
–20 –0.50 dB (47)________ 0.50 dB ±0.078 dB
–24 –0.54 dB (48)________ 0.54 dB ±0.074 dB
–28 –0.58 dB (49)________ 0.58 dB ±0.073 dB
–32 –0.62 dB (50)________ 0.62 dB ±0.077 dB
–36 –0.66 dB (51)________ 0.66 dB ±0.075 dB
–40 –0.70 dB (52)________ 0.70 dB ±0.081 dB
–44 –0.74 dB (53)________ 0.74 dB ±0.077 dB
–48 –0.78 dB (54)________ 0.78 dB ±0.076 dB
–52 –0.82 dB (55)________ 0.82 dB ±0.080 dB
–56 –0.86 dB (56)________ 0.86 dB ±0.078 dB
–60 –0.90 dB (57)________ 0.90 dB ±0.084 dB
–64 –0.94 dB (58)________ 0.94 dB ±0.081 dB
–68 –0.98 dB (59)________ 0.98 dB ±0.080 dB
–72 –1.02 dB (60)________ 1.02 dB ±0.084 dB
–76 –1.06 dB (61)________ 1.06 dB ±0.083 dB
–80 –1.10 dB (62)________ 1.10 dB ±0.089 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty