Chapter 3 521
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
65 ms Gate Length 64.993 ms (4)________ 65.007 ms ±647 ns
51. Gate Mode Additional
Amplitude Error (Option
1D6 only)
Amplitude Error –0.2 dB (1)________ 0.2 dB ±0.023 dB
58. GSM Phase and
Frequency Error
(Option BAH and B7E)
Peak Phase Error –2.1 Deg (1)________ 2.1 Deg 0.52 Deg
RMS Phase Error –1.1 Deg (2)________ 1.1 Deg 0.000063 Deg
Frequency Error –10 Hz (3)________ 10 Hz ±0.006 Hz
59. Comms Absolute Power
Accuracy
(Options BAC or BAH)
20 to 30° C
cdmaOne Channel Power
Accuracy
(Option BAC only)
Cellular Band
Input Amplitude
15 dBm –0.82 dB (1)________ 0.82 dB ±0.173 dB
−5 dBm –0.82 dB (2)________ 0.82 dB ±0.161 dB
−25 dBm –0.78 dB (3)________ 0.78 dB ±0.161 dB
−45 dBm –0.69 dB (4)________ 0.69 dB ±0.152 dB
−55 dBm –0.77 dB (5)________ 0.77 dB ±0.152 dB
−70 dBm –0.89 dB (6)________ 0.89 dB ±0.161 dB
PCS Band
Input Amplitude
15 dBm –0.78 dB (7)________ 0.78 dB ±0.173 dB
−5 dBm –0.78 dB (8)________ 0.78 dB ±0.161 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty