Chapter 3 443
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
25. Frequency Response
(Preamp On)
(Option 1DS only)
Note: Enter data in the appropriate section below depending upon the
ambient temperature at which the test was performed and installed
options.
20 to 30° C
50 Ω Input (No Option
1DP)
(100 kHz to 1.5 GHz)
Maximum Response (1)________ 1.0 dB ±0.28 dB
Minimum Response –1.0 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 2.0 dB ±0.28 dB
Option 1DP (75 Ω)
(1 MHz to 1.5 GHz)
Maximum Response (1)________ 1.50 dB ±0.28 dB
Minimum Response –1.50 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 3.0 dB ±0.28 dB
0 to 55° C
50 Ω Input (No Option
1DP)
(100 kHz to 1.5 GHz)
Maximum Response (1)________ 1.5 dB ±0.28 dB
Minimum Response –1.5 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 3.0 dB ±0.28 dB
Option 1DP (75 Ω)
(1 MHz to 1.5 GHz)
Maximum Response (1)________ 2.0 dB ±0.28 dB
Minimum Response –2.0 dB (2)________ ±0.28 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty